Twenty-four-hour Bence-Jones protein determinations: can we ensure accuracy?
Jennifer S Kaplan1, Gary L Horowitz
1Department of Pathology, Beth Israel Deaconess Medical Center, Boston, Massachusetts 02215, USA.
Context:
Light chain disease represents 15% to 20% of cases of multiple myeloma. Current guidelines recommend monitoring these patients with 24-hour urine collections.
Objective:
To determine the reliability of 24-hour urine collections in assessing the amount of Bence-Jones protein (BJP).
Design:
We included all patients from our institution from 2003 through 2008 with BJP who had more than four 24-hour urine collections. We compared BJP excretion calculated from the submitted 24-hour collection with BJP excretion calculated by normalizing the collection to that patient's mean 24-hour creatinine excretion. We also looked at differences in serial values with these 2 methods. In addition, we evaluated the feasibility of using random urine samples to determine BJP excretion.
Results:
A total of 14 patients with 135 24-hour urine collections met our inclusion criteria. The 24-hour urine creatinine excretion for each patient, which should be reasonably constant, varied considerably (coefficient of variation range 12%-30%). Differences in the 2 methods of calculating BJP excretion ranged from -1588 to 2315 mg/d. Among a total of 121 serial 24-hour measurements, the differences were clinically significant in 37 (30%). Among a total of 23 random urine samples from 11 of these patients submitted within 10 days of a 24-hour collection, the estimated BJP excretion appeared to be accurate in at least 18 (78%).
Conclusions:
Twenty-four-hour urine collections for BJP are, in practice, often misleading. At a minimum, one should verify that the 24-hour creatinine excretion is accurate. In addition, it may be possible to use the protein/creatinine ratio from random urine samples to determine 24-hour BJP excretion.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...