Related Experiment Video
Updated: May 30, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Semiconductor detectors allow low-dose-low-dose 1-day SPECT myocardial perfusion imaging
Rene Nkoulou1, Aju P Pazhenkottil, Silke M Kuest
1Deparment of Radiology, Cardiac Imaging, University Hospital Zurich, Zurich, Switzerland.
Unlabelled:
Cadmium zinc telluride (CZT) detectors with linear counting rate response enable count subtraction in sequential scanning. We evaluated whether count subtraction eliminated the need for higher activity doses in the second part of the 1-d stress-rest myocardial perfusion imaging (MPI) protocol.
Methods:
For 50 patients (mean age ± SD, 66 ± 12 y) with visually abnormal (n = 42) or equivocal (n = 8) adenosine-stress MPI (320 MBq of (99m)Tc-tetrofosmin) on a CZT camera, rest MPI was performed with a low dose (320 MBq) and repeated after injection of an additional 640 MBq of (99m)Tc-tetrofosmin to achieve a standard 3-fold increased dose at rest (960 MBq), compared with stress (320 MBq). Low-dose rest myocardial perfusion images were reconstructed after subtracting the background activity of the preceding stress scan. Segmental percentage tracer uptake of the 2 rest myocardial perfusion images (320 vs. 960 MBq) was compared using intraclass correlation and Bland-Altman limits of agreement. Patient- and coronary territory-based clinical agreement was assessed.
Results:
The standard protocol revealed ischemia in 34 (68%) and a fixed defect in 8 (16%) patients, of whom 33 (97%) and 8 (100%) were correctly identified by low-dose MPI (clinical agreement, 98%). Segmental uptake correlated well between low- and standard-dose rest scans (r = 0.94, P < 0.001; Bland-Altman limits of agreement, -11 to +11%). Defect extent was 14.4% (low-dose) versus 13.1% (standard-dose) at rest (P = not statistically significant) and 26.6% at stress (P < 0.001 vs. rest scans).
Conclusion:
These promising results suggest that accurate assessment of ischemic myocardial disease is feasible with a low-dose-low-dose 1-d SPECT MPI protocol using a CZT device.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...