Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Effect of detector lag on CT noise power spectra
1Department of Radiology, Stanford University, Stanford, California 94305, USA. bjd1219@stanford.edu
Purpose:
The authors examined the effect of detector lag on the noise power spectrum (NPS) of CT images reconstructed with filtered backprojection (FBP).
Methods:
The authors derived an analytical expression of the NPS with detector lag, and then verified it using computer simulations with parallel beam and fan beam geometries. The dependence of the NPS on the amount of lag, location within the scanned field of view (FOV), and the number of views used in the reconstruction (samples per rotation) was investigated using constant and view dependent noise in the raw data.
Results:
Detector lag introduces noise correlation in the azimuthal direction. The effect on the NPS is a frequency dependent reduction in amplitude. In small regions of the image, the effect is primarily in the frequencies corresponding to the azimuthal direction. The noise blurring and NPS filtering increases with increasing radial distance, and therefore regions at larger radial distances have lower noise power. With the same detector lag response function, the amount of noise correlation and NPS filtering decreases with increasing number of views.
Conclusions:
The shape of the NPS depends on the detector lag coefficients, location of the region, and the number of views used in the reconstruction. In general, the noise correlation caused by detector lag decreased the amplitude of the NPS.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
NMR Spectrometers: Resolution and Error Correction
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)