Dose reduction using bismuth shielding during paediatric CT examinations in Slovakia
L' Gbelcová1, D Nikodemová, M Horváthová
1Faculty of Health Care and Social Work, Trnava University in Trnava, Univerzitné nám 1, 918 43 Trnava, Slovakia. ubikag@gmail.com
Bismuth shielding significantly reduces radiation dose in pediatric head CT scans, with one department showing a 56-65% reduction. This protective measure is crucial for minimizing radiation exposure in children undergoing computed tomography (CT) examinations.
Area of Science:
- Medical Physics
- Radiology
- Pediatric Imaging
Background:
- Increasing use of computed tomography (CT) examinations in Slovakia raises concerns about population radiation exposure.
- Children are more susceptible to radiation effects than adults, necessitating dose monitoring in pediatric patients.
Purpose of the Study:
- To assess radiation doses in pediatric patients undergoing CT scans in Slovakia.
- To evaluate the effectiveness of bismuth shielding in reducing eye and thyroid doses during pediatric head CT examinations.
Main Methods:
- Utilized an anthropomorphic child phantom for radiation dose measurements.
- Exposed 215 thermoluminescent dosimeters on the phantom with and without bismuth shields.
- Conducted measurements across three selected Slovakian hospital departments.
Main Results:
- Two of the three departments showed a reduction in radiation dose with bismuth shields.
- One department reported a significant dose reduction ranging from 56% to 65% for critical organs.
- Effectiveness of shielding varied between the examined departments.
Conclusions:
- Bismuth shielding is a potentially effective method for reducing radiation dose to the eyes and thyroid in pediatric head CT scans.
- The significant dose reduction observed in one department highlights the importance of implementing and optimizing shielding protocols.
- Further studies are needed to confirm widespread efficacy and establish standardized guidelines for bismuth shielding in pediatric CT.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Pharmacokinetics in Pediatric Patients: Drug Excretion
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET