Related Experiment Video
Updated: Apr 25, 2026

Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
Acquisition of MV-scatter-free kilovoltage CBCT images during RapidArc™ or VMAT
Clifton Ling1, Pengpeng Zhang, Thanos Etmektzoglou
1Varian Medical Systems, Memorial Sloan Kettering Cancer Center, USA. lingc@mskcc.org
Purpose:
To perform kilovoltage (kV) cone beam computed tomography (CBCT) imaging concomitant with the delivery of megavoltage (MV) RapidArc treatment, and demonstrate the feasibility of obtaining MV-scatter-free kV CBCT images.
Methods And Materials:
RapidArc/CBCT treatment and imaging plans are designed, and delivered on the Varian TrueBeam, using its Developer Mode. The plan contains 250 control points for MV-radiation delivery, each over an arc of 0.4-0.7(o). Interlaced between successive MV delivery control points are imaging control points, each over an arc of 0.7-1.1(o). During the 360(o) gantry rotation for the RapidArc delivery, CBCT projections of a phantom are acquired at 11 frames per second. The kV projections with minimal MV-scatter are selected, based on gantry angle, and the CBCT(s) image reconstructed. For comparison, a reference CBCT(r) image is acquired in the normal way. In addition, to examine the effect of MV-scatter we acquire CBCT(c) using the same treatment plan without the imaging control points, i.e. with continuous MV delivery during the 360(o) rotation. Quantitative evaluation of image qualities is performed based on the concepts of CNR (contrast-to-noise ratio) and NSTD (normalized standard deviation).
Results:
The different types of CBCT images were reconstructed, evaluated, and compared. Visual comparison indicates that the image quality of CBCT(s) is similar to that of the reference CBCT(r), and that the quality of CBCT(c) is significantly degraded by the MV-scatter. Quantitative evaluation of the image quality indicates that MV-scatter significantly decreases the CNR of CBCT (from ∼7 to ∼3.5 in one comparison). Similarly, MV-scatter significantly increases the inhomogeneity of image intensity, e.g. from ∼0.03 to ∼0.06 in one comparison.
Conclusion:
We have developed a method to acquire MV-scatter-free kV CBCT images concomitant with the delivery of RapidArc treatment. Engineering development is necessary to improve the process, e.g. by synchronization of the MV and kV beams.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography

