Development and Angiographic Use of the Rabbit VX2 Model for Liver Cancer
Published on: January 7, 2019
Treatment of rabbit liver cancer in vivo using miniaturized image-ablate ultrasound arrays
T Douglas Mast1, Peter G Barthe, Inder Raj S Makin
1Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH 45267-0586, USA. doug.mast@uc.edu
Abstract:
In the preclinical studies reported here, VX2 cancer within rabbit liver has been treated by bulk ultrasound ablation employing miniaturized image-ablate arrays. Array probes were constructed with 32 elements in a 2.3 × 20 mm(2) aperture, packaged within a 3.1 mm stainless steel tube with a cooling and coupling balloon for in vivo use. The probes were measured capable of 50% fractional bandwidth for pulse-echo imaging (center frequency 4.4 MHz) with >110 W/cm(2) surface intensity available at sonication frequencies 3.5 and 4.8 MHz. B-scan imaging performance of the arrays was measured to be comparable to larger diagnostic linear arrays, although nearfield image quality was reduced by ringdown artifacts. A series of in vivo ablation procedures was performed using an unfocused 32-element aperture firing at 4.8 MHz with exposure durations 20-70.5 s and in situ spatial average, temporal average intensities 22.4-38.5 W/cm(2). Ablation of a complete tumor cross-section was confirmed by vital staining in seven of 12 exposures, with four exposures ablating an additional margin >1 mm beyond the tumor in all directions. Analysis suggests a threshold ablation effect, with complete ablation of tumor cross-sections for exposures with delivery of >838 J acoustic energy. The results show feasibility for in vivo liver cancer ablation using miniaturized image-ablate arrays suitable for interstitial deployment.