Comparison of fully automated and semi-automated biopsy needles for lung biopsy under CT fluoroscopic guidance
R Yoshimatsu1, T Yamagami, O Tanaka
1Department of Radiology, Graduate School of Medical Science, Kyoto Prefectural University of Medicine, Kyoto, Japan. rika442@koto.kpu-m.ac.jp
Objective:
The aim of this study was to compare two different automated biopsy needles, a fully automated biopsy needle (Monopty; Bard, Covington, GA) and a semi-automated biopsy needle (Temno; Bauer Medical, Clearwater, FL), for lung biopsy.
Methods:
50 consecutive percutaneous lung biopsies using the Monopty needle between June 2006 and January 2007 and 66 consecutive lung biopsies for 1 nodule in each session using the Temno needle between February 2007 and August 2008 were performed under CT fluoroscopic guidance followed by histopathological evaluation.
Results:
In 42/50 lung biopsies performed with the Monopty needle and 54/66 lung biopsies performed with the Temno needle, the final diagnosis was confirmed by independent surgical pathological findings or clinical follow-up. Sufficient samples for histopathological evaluation were obtained in all 50 (100%) biopsies using the Monopty needle and in 55 (83.3%) of the 66 biopsies using the Temno needle (p<0.01). Accurate diagnosis was achieved in 41 (97.6%) of 42 biopsies using the Monopty needle and in 45 (83.3%) of 54 biopsies using the Temno needle (p=0.04). Biopsy-induced complications were pneumothorax, haemoptysis and haemothorax in 44.0%, 10.0% and 6.0% of biopsies, respectively, using the Monopty needle and in 48.3%, 8.3% and 3.3%, respectively, using the Temno needle.
Conclusion:
There is a possibility that a fully automated biopsy needle such as the Monopty is more useful for CT scan-guided lung biopsy than semi-automated biopsy needles.