Objective:
To prospectively evaluate the accuracy of left and right ventricular function and myocardial mass measurements based on a dual-step, low radiation dose protocol with prospectively ECG-triggered 2nd generation dual-source CT (DSCT), using cardiac MRI (cMRI) as the reference standard.
Materials And Methods:
Twenty patients underwent 1.5T cMRI and prospectively ECG-triggered dual-step pulsing cardiac DSCT. This image acquisition mode performs low-radiation (20% tube current) imaging over the majority of the cardiac cycle and applies full radiation only during a single adjustable phase. Full-radiation-phase images were used to assess cardiac morphology, while low-radiation-phase images were used to measure left and right ventricular function and mass. Quantitative CT measurements based on contiguous multiphase short-axis reconstructions from the axial CT data were compared with short-axis SSFP cardiac cine MRI. Contours were manually traced around the ventricular borders for calculation of left and right ventricular end-diastolic volume, end-systolic volume, stroke volume, ejection fraction and myocardial mass for both modalities. Statistical methods included independent t-tests, the Mann-Whitney U test, Pearson correlation statistics, and Bland-Altman analysis.
Results:
All CT measurements of left and right ventricular function and mass correlated well with those from cMRI: for left/right end-diastolic volume r=0.885/0.801, left/right end-systolic volume r=0.947/0.879, left/right stroke volume r=0.620/0.697, left/right ejection fraction r=0.869/0.751, and left/right myocardial mass r=0.959/0.702. Mean radiation dose was 6.2±1.8mSv.
Conclusions:
Prospectively ECG-triggered, dual-step pulsing cardiac DSCT accurately quantifies left and right ventricular function and myocardial mass in comparison with cMRI with substantially lower radiation exposure than reported for traditional retrospective ECG-gating.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Magnetic Resonance Imaging