Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Accuracy assessment of the axial images obtained from cone beam computed tomography
Francine Kühl Panzarella1, J L C Junqueira, L B Oliveira
1Faculdade São Leopoldo Mandic, School of Dentistry, Campinas, Brazil. prof.francine@slmandic.edu.br
Objective:
The aim of this study was to evaluate accuracy of linear measurements assessed from axial tomograms and the influence of the use of different protocols in two cone beam CT (CBCT) units.
Methods:
A cylinder object consisting of Nylon® (Day Brazil, Sao Paulo, Brazil) with radiopaque markers was radiographically examined applying different protocols from NewTom 3G(TM) (Quantitative Radiology s.r.l, Verona, Veneto, Italy) and i-CAT(TM) (Imaging Sciences International, Hatfield, PA) units. Horizontal (A-B) and vertical (C-D) distances were assessed from axial tomograms and measured using a digital calliper that provided the gold standard for actual values.
Results:
There were differences when considering acquisition protocols to each CBCT unit. Concerning all analysed protocols from i-CAT(TM) and Newtom 3G(TM), both A-B and C-D distances presented underestimated values. Measurements of the axial images obtained from NewTom 3G(TM) (6 inch 0.16 mm and 9 inch 0.25 mm) were similar to the ones obtained from i-CAT(TM) (13 cm 20 s 0.3 mm, 13 cm 20 s 0.4 mm and 13 cm 40 s 0.25 mm).
Conclusion:
The use of different protocols from CBCT machines influences linear measurements assessed from axial images. Linear distances were underestimated in both equipments. Our findings suggest that the best protocol for the i-CAT(TM) is 13 cm 20 s 0.3 mm and for the NewTom 3G(TM), the use of 6 inch or 9 inch is recommended.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...