Electrical impedance tomography measured at two thoracic levels can visualize the ventilation distribution changes at
Ido G Bikker1, Carsten Preis, Mahamud Egal
1Department of Intensive Care Medicine, Erasmus MC, 's-Gravendijkwal 230, NL-3015GE Rotterdam, The Netherlands.
Introduction:
Computed tomography of the lung has shown that ventilation shifts from dependent to nondependent lung regions. In this study, we investigated whether, at the bedside, electrical impedance tomography (EIT) at the cranial and caudal thoracic levels can be used to visualize changes in ventilation distribution during a decremental positive end-expiratory pressure (PEEP) trial and the relation of these changes to global compliance in mechanically ventilated patients.
Methods:
Ventilation distribution was calculated on the basis of EIT results from 12 mechanically ventilated patients after cardiac surgery at a cardiothoracic ICU. Measurements were taken at four PEEP levels (15, 10, 5 and 0 cm H2O) at both the cranial and caudal lung levels, which were divided into four ventral-to-dorsal regions. Regional compliance was calculated using impedance and driving pressure data.
Results:
We found that tidal impedance variation divided by tidal volume significantly decreased on caudal EIT slices, whereas this measurement increased on the cranial EIT slices. The dorsal-to-ventral impedance distribution, expressed according to the center of gravity index, decreased during the decremental PEEP trial at both EIT levels. Optimal regional compliance differed at different PEEP levels: 10 and 5 cm H2O at the cranial level and 15 and 10 cm H2O at the caudal level for the dependent and nondependent lung regions, respectively.
Conclusions:
At the bedside, EIT measured at two thoracic levels showed different behavior between the caudal and cranial lung levels during a decremental PEEP trial. These results indicate that there is probably no single optimal PEEP level for all lung regions.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Atelectasis II: Pathophysiology
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...