Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
Published on: January 7, 2016
Quantification of human growth hormone by amino acid composition analysis using isotope dilution
Ji-Seon Jeong1, Hyuk-Min Lim, Sook-Kyung Kim
1Center for Bioanalysis, Department of Metrology for Quality of Life, Korea Research Institute of Standards and Science, 1 Doryong-dong, Yuseong-gu, Daejeon 305-340, South Korea.
Abstract:
We describe an accurate method for protein quantification based on conventional acid hydrolysis and an isotope dilution-HPLC-mass spectrometry (ID-HPLC-MS) method. Sample purity was confirmed using capillary zone electrophoresis, HPLC and MS. The analyte protein, human growth hormone (hGH), was effectively hydrolyzed by incubation with 8 M hydrochloric acid at 130 °C for 48 h, where at least 1 μM of hGH was treated to avoid possible degradation of released amino acids during hydrolysis. Using a reversed-phase column, the analytes (isoleucine, phenylalanine, proline and valine) were separated within 5 min using an isocratic eluent comprising 10% acetonitrile containing 0.1% trifluoroacetic acid. The detection limit (signal to noise ratio of 3) of amino acids was 5.5-6.2 fmol per injection. The quantification precision (RSD) of amino acids for intra- and inter-day assays was less than 0.98% and 0.39%, respectively. Comparison with other biochemical and instrumental methods revealed substantially higher accuracy and reproducibility of the ID-HPLC-MS/MS method as expected. The optimized hydrolysis and analytical conditions in our study were suitable for accurate quantification of hGH.