An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Load fatigue performance of conical implant-abutment connections
Y L Seetoh1, Keson B Tan, E K Chua
1Department of Restorative Dentistry, National Dental Centre, Republic of Singapore.
Purpose:
Conical implant-abutment connections for platform switching have been recently introduced in implant systems. This study investigated the load fatigue performance of three conical abutment systems and their corresponding titanium and zirconia abutments.
Materials And Methods:
Regular-diameter implants of the Ankylos (AK), PrimaConnex (PC), and Straumann (ST) systems were tested with their corresponding titanium (Ti) and zirconia (Zr) abutments tightened to the recommended torque (n = 5 implant-abutment assemblies per group). A rotational load fatigue machine applied a sinusoidally varying tensile-compressive 21 N load to specimens at a 45-degree angle, producing an effective bending moment of 35 Ncm at a frequency of 10 Hz. The number of cycles to failure was recorded, with the upper limit set at 5 million cycles. Results were evaluated through analyses of variance.
Results:
Except for the ST Zr group, which showed no failures in four samples and one failure just below the screw head, and the AK Ti group, in which one sample was preserved without fracture, all groups experienced failure of at least one of the components, whether the abutment screw only, the abutment, and/or the implant neck. There were significant differences between systems. There was no difference between systems for the Ti abutments, and the ST group was significantly different from the AK and PC groups for the Zr abutments.
Conclusions:
Ti conical abutments appear to have poorer load fatigue performance compared with earlier studies of external-hexagon connections. The load fatigue performance of Zr conical abutments varied and seemed to be highly system dependent. Many of the fractures in both the Ti and Zr abutment groups occurred within the implant, and retrieval would pose a significant clinical challenge. The clinician should weigh the mechanical, biologic, and esthetic considerations before selection of any implant system, connection type, or abutment material.
Related Concept Videos
Fatigue
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Normal Strain under Axial Loading
Bearing Stress
Due to the intricacy of these microforces, an average value, known as bearing stress, is often used by...
Internal Loadings in Structural Members: Problem Solving
To illustrate this, let's consider a beam OC of 5 kN, inclined at an angle of 53.13° with the horizontal and supported at both ends. Determine the internal loadings...