Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challenge
P Z Y Liu1, N Suchowerska, J Lambert
1School of Physics, The University of Sydney, NSW 2006, Australia. pliu@physics.usyd.edu.au
Abstract:
In scintillation dosimetry, a Cerenkov background signal is generated when a conventional fibre optic is exposed to radiation produced by a megavoltage linear accelerator. Three methods of measuring dose in the presence of Cerenkov background are compared. In the first method, a second background fibre is used to estimate the Cerenkov signal in the signal fibre. In the second method, a colour camera is used to measure the combined scintillation and Cerenkov light in two wavelength ranges and a mathematical process is used to extract the scintillation signal. In the third method, a hollow air core light guide is used to carry the scintillation signal through the primary radiation field. In this paper, the strengths and weaknesses of each dosimetry system are identified and recommendations for the optimum method for common clinical dosimetry situations are made.