Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
A practical approach to estimate the weighted CT dose index over an infinite integration length
1Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, Boston, MA 02114, USA.
Abstract:
The CT scanner-displayed radiation dose information is based on CT dose index (CTDI) over an integration length of 100 mm (CTDI(100)), which is lower than the CTDI over an infinite integration length (CTDI(∞)). In an adult or a pediatric body CT scan, the limiting equilibrium dose can be established near the central scan plane, and CTDI(∞) more closely indicates the accumulated dose than CTDI(100). The aim of this study was to (a) evaluate CTDI efficiencies, ϵ(CTDI(100)) = CTDI(100)/CTDI(∞), for a multi-detector CT (MDCT) scanner, (b) examine the dependences of ϵ(CTDI(100)) on kV, beam width, phantom diameter, phantom length and position in phantom and (c) investigate how to estimate CTDI(∞) based on the CT scanner-displayed information. We performed a comprehensive Geant4-based simulation study of a clinical CT scanner, and calculated ϵ(CTDI(100)) for a range of parameters. The results were compared with the ϵ(CTDI(100)) data of previous studies. Differences in the ϵ(CTDI(100)) values of these studies were assessed. A broad analysis of the ϵ(CTDI(100)) variations with the above-mentioned parameters was presented. Based on the results, we proposed a practical approach to obtain the weighted CTDI(∞) using the CT scanner-displayed information. A reference combination of 120 kV and a beam width close to 20 mm can be selected to determine the efficiencies of the weighted CTDI by using either phantom measurements or computer simulations. The results can be applied to estimate the weighted CTDI(∞) for 80-140 kV and the beam widths within 40 mm. Errors in the weighted CTDI(∞) due to the variations of kV and beam width can be 5% or less for the MDCT scanners.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:00Spatial Measurements of Perfusion, Interstitial Fluid Pressure and Liposomes Accumulation in Solid Tumors
Published on: August 18, 2016
Related Concept Videos
Improper Integrals: Infinite Intervals
Approximate Integration
Definite Integral
Improper Integrals: Discontinuous Integrands
Fundamental Theorem of Calculus I
Biological Effects of Radiation