Usefulness of specific calibration coefficients for gamma-emitting sources measured by radionuclide calibrators in nuclear medicine
Related Experiment Video
Updated: May 30, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
François O Bochud1, Jean-Pascal Laedermann, Sébastien Baechler
1Institute of Radiation Physics, University Hospital and University of Lausanne, Rue du Grand-Prd 1, CH-1007 Lausanne, Switzerland. francois.bochud@chuv.ch
Purpose:
In nuclear medicine, the activity of a radionuclide is measured with a radionuclide calibrator that often has a calibration coefficient independent of the container type and filling.
Methods:
To determine the effect of the container on the accuracy of measuring the activity injected into a patient, The authors simulated a commercial radionuclide calibrator and 18 container types most typically used in clinical practice. The instrument sensitivity was computed for various container thicknesses and filling levels. Monoenergetic photons and electrons as well as seven common radionuclides were considered.
Results:
The quality of the simulation with gamma-emitting sources was validated by an agreement with measurements better than 4% in five selected radionuclides. The results show that the measured activity can vary by more than a factor of 2 depending on the type of container. The filling level and the thickness of the container wall only have a marginal effect for radionuclides of high energy but could induce differences up to 4%.
Conclusions:
The authors conclude that radionuclide calibrators should be tailored to the uncertainty required by clinical applications. For most clinical cases, and at least for the low-energy gamma and x-ray emitters, measurements should be performed with calibration coefficients specific to the container type.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Biological Effects of Radiation
Calibration Curves: Correlation Coefficient
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...