Related Experiment Video
Updated: May 30, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
The effect of tracheal tube size on air leak around the cuffs
Jin-Young Hwang1, Sang-Hyun Park, Sung-Hee Han
1Department of Anesthesiology and Pain Medicine, Seoul National University Bundang Hospital, Seongnam, Korea.
Background:
This randomized single-blinded, cross-over study was done to evaluate the influence of the size of tracheal tubes on air leaks around the cuffs.
Methods:
In a benchtop model, the number of longitudinal folds on the cuffs was evaluated for different sizes of tracheal tubes. In an anesthetized patient study, thirty patients scheduled for elective surgery under general anesthesia were included. After induction of anesthesia, the trachea was intubated with two sizes of tracheal tubes in a random sequence: in men, internal diameter of 7.5 mm and 8.0 mm; in women, internal diameter of 7.0 mm and 7.5 mm. After tracheal intubation with each tube, air leak pressures were evaluated at intracuff pressures of 20, 25 and 30 cmH(2)O by auscultation. To calculate the tracheal tube resistance (R), an inspiratory pause of 20% was applied and the resulting peak airway pressure (P(peak)), plateau pressure (P(pl)) and mean expiratory tidal volume (Flow) were inserted in the formula R = (P(peak) - P(pl))/Flow.
Results:
More longitudinal folds of the tracheal tube cuffs occurred in larger sized tubes compared to the smaller ones in a benchtop model. Air leakage was significantly less for the smaller tracheal tubes than for the larger ones for each gender at intracuff pressures of 20, 25 and 30 cmH(2)O. Tracheal tube resistances were not significantly altered by the size of tracheal tube.
Conclusions:
The use of a smaller tracheal tube within an acceptable size can reduce air leakage around the cuff without significantly changing the tracheal tube resistance.
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...