Objective:
The purpose of this study was to evaluate the impact of the use of 64-MDCT and volume image reading on the radiologic workflow during a mass casualty incident simulation.
Materials And Methods:
For this simulation, casualties were taken to our level I trauma center, and triage was done with whole-body 64-MDCT. The complete raw dataset of thin-section images was sent to a dedicated 3D workstation for further interpretation and simultaneous reformations. This new reading method is called volume image reading. Several time frames were documented to evaluate the workflow: examination time, time needed for image processing, and mean image transfer rates. The results were compared with those of a previous study using a 4-MDCT with axial images only and transfer of data to a PACS.
Results:
The time for complete image processing (acquisition, reconstruction, and transfer) for 64-MDCT was 4.1 minutes (range, 3.9-4.3 minutes) compared with 9.0 minutes (range, 6.4-10.2 minutes) for 4-MDCT (p ≤ 0.001). The image processing capacity was 14.8 examinations/h for 64-MDCT compared with 6.7 examinations/h for 4-MDCT. The mean number of images was 953 for 64-MDCT compared with 202 for 4-MDCT (p ≤ 0.001). There were no significant differences between 64- and 4-MDCT for the time needed to prepare patients.
Conclusion:
The use of 64-MDCT with volume image reading led to evident advantages in the radiologic trauma workflow compared with 4-MDCT. Reading of the full image set including reformations can be initiated earlier with volume image reading.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography