Optimizing the driving scheme of a self-actuated atomic force microscope probe for high-speed applications
Related Experiment Video
Updated: May 29, 2026

Active Probe Atomic Force Microscopy with Quattro-Parallel Cantilever Arrays for High-Throughput Large-Scale Sample Inspection
Published on: June 13, 2023
1Department of Electrical and Electronics Engineering, Bahçeşehir University, Beşiktaş, İstanbul 34353, Turkey. mujdat.balantekin@bahcesehir.edu.tr
Abstract:
We investigate the optimum driving scheme of a dynamic atomic force microscope with a self-actuated probe for high-speed applications by performing numerical simulations. We compare the recently developed methods such as Q-control, dynamic PID control, and modified Q-control methods to the standard tapping mode by considering scan speed and peak transient forces. In addition, the effects of driving frequency and set-point amplitude on the maximum achievable scan speed for the same probe-sample system are discussed. We find that the scan speed can be increased significantly at the expense of increased peak transient forces.