Effects of high blood flow and high pre-dilution replacement fluid rates on small solute clearances in hemofiltration
Casey N Gashti1, Roger A Rodby, Zhongping Huang
1Section of Nephrology, Rush University Medical Center, Chicago, Ill., USA. Casey_Gashti @ rush.edu
Background/Aims:
In pre-dilution hemofiltration (HF), solute clearance is less than the HF rate. While the amount of this loss is predictable, it has not been validated in high-volume HF associated with high blood flow rates.
Methods:
Using isovolemic pre-dilution HF, we studied small solute clearances using combinations of blood flow (Q(B); 150, 250, 350, 450 ml/min) and replacement fluid (RF) flow (Q(RF); 2, 4, 6 l/h) to determine clearance losses we entitled 'measured efficiency' (E(M)). E(M) was compared to predicted efficiency (E(P)) = (Q(B)/Q(B) + Q(RF)).
Results:
Pre-dilution produced E(M) values of 61-93%. Increases in Q(B) for any Q(RF) and decreases in Q(RF) for any Q(B) increased E(M) over a wide range of Q(B) and Q(RF). E(P) was equivalent to E(M).
Conclusion:
In high-volume pre-dilution HF, E(P) can be used to determine E(M) across a broad range of Q(B) and Q(RF) values. Higher Q(RF) requires higher Q(B) to minimize the attenuating effects of pre-dilution on clearance.
Related Concept Videos
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Hemodialysis II: Procedure and Complications
Continuous Renal Replacement Therapy
Formation of Dilute Urine
Filtrate Osmolarity in the PCT
Initially, as the filtrate passes through the proximal convoluted tubule (PCT), its...
Fluid Movement Between Compartments