Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Imaging Studies III: Computed Tomography01:27

Imaging Studies III: Computed Tomography

389
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
389
Computed Tomography01:10

Computed Tomography

8.8K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
8.8K
Electrocardiogram01:29

Electrocardiogram

6.3K
An electrocardiogram (ECG or EKG) is a critical diagnostic tool that records the electrical signals produced by the heart during each heartbeat. This recording is achieved through electrodes placed strategically on the arms, legs, and chest. The electrocardiograph amplifies these signals and produces 12 distinct tracings, offering a comprehensive understanding of the heart's electrical activity.
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
6.3K
Electron Microscope Tomography and Single-particle Reconstruction01:07

Electron Microscope Tomography and Single-particle Reconstruction

2.9K
Transmission electron microscopy (TEM) can be used to determine the 3D structure of biological samples with the help of techniques such as electron microscope tomography and single-particle reconstruction. While single-particle reconstruction can examine macromolecules and macromolecular complexes in vitro conditions only, tomography permits the study of cell components or small cells in vivo.
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
2.9K
Electrocardiogram Fundamentals01:28

Electrocardiogram Fundamentals

1.6K
Introduction
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
1.6K
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches01:23

Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches

475
Biopharmaceutical studies constitute a vital field aiming to enhance drug delivery methods and refine therapeutic approaches, drawing upon diverse interdisciplinary knowledge. In research methodologies, the choice between controlled and non-controlled studies significantly influences the study's reliability and accuracy.
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
475

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Soft tissue-based selection of alignment and implant strategy in total knee arthroplasty is associated with comparable patient-reported outcomes despite baseline differences.

The Knee·2026
Same author

Technical Note: Efficacy of CT spatial resolution metrics in clinical settings.

Medical physics·2026
Same author

[3. Metal Artifact Reduction in X-ray CT].

Nihon Hoshasen Gijutsu Gakkai zasshi·2026
Same author

Selection of Radiological Physics and Technology Awards 2025.

Radiological physics and technology·2026
Same author

Megavoltage computed tomography image quality assessment in a modern tomotherapy system: a frequency-domain analysis.

Radiological physics and technology·2026
Same author

Improvement of blooming artifact in coronary CT image using high-resolution kernel and image-based noise reduction.

Radiological physics and technology·2025

Related Experiment Video

Updated: Feb 8, 2026

Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography
04:48

Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography

Published on: November 30, 2022

3.5K

[A comparative study of electrocardiogram multi-segment reconstruction and dual source computed tomography using a

Kazuya Ohashi1, Katsuhiro Ichikawa, Ryo Higashide

  • 1Central Department of Radiology, Nagoya City University Hospital.

Nihon Hoshasen Gijutsu Gakkai Zasshi
|August 27, 2011
PubMed
Summary

Dual Source CT (DSCT) offers superior image quality for coronary CT angiography compared to electrocardiogram-gated multi-segment reconstruction (EMR). DSCT effectively reduces motion artifacts, enhancing diagnostic accuracy in cardiac imaging.

More Related Videos

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
04:40

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans

Published on: August 28, 2018

16.1K
Dorsal Column Steerability with Dual Parallel Leads using Dedicated Power Sources: A Computational Model
11:19

Dorsal Column Steerability with Dual Parallel Leads using Dedicated Power Sources: A Computational Model

Published on: February 10, 2011

12.3K

Related Experiment Videos

Last Updated: Feb 8, 2026

Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography
04:48

Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography

Published on: November 30, 2022

3.5K
Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
04:40

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans

Published on: August 28, 2018

16.1K
Dorsal Column Steerability with Dual Parallel Leads using Dedicated Power Sources: A Computational Model
11:19

Dorsal Column Steerability with Dual Parallel Leads using Dedicated Power Sources: A Computational Model

Published on: February 10, 2011

12.3K

Area of Science:

  • Medical Imaging
  • Cardiovascular Technology
  • Radiology

Background:

  • Improving temporal resolution in coronary CT is crucial for reducing motion artifacts.
  • Electrocardiogram-gated multi-segment reconstruction (EMR) and dual-source CT (DSCT) are primary methods for enhancing temporal resolution.

Purpose of the Study:

  • To evaluate and compare the image quality of EMR and DSCT for cardiac CT.
  • To assess the effectiveness of a novel motion phantom system in evaluating cardiac CT techniques.

Main Methods:

  • Development of a motion phantom system simulating cyclic motion at variable speeds.
  • Acquisition of coronary CT images using the phantom with a coronary artery-like acrylic rod.
  • Evaluation of motion artifacts by analyzing image profiles and pixel displacement for EMR (16-row multi-slice CT) and DSCT.

Main Results:

  • EMR, when implemented on 16-row multi-slice CT, resulted in significant motion artifacts and displacement.
  • DSCT demonstrated superior performance, producing excellent images with substantially fewer motion artifacts.
  • The motion phantom system effectively validated the differences in image quality between the two CT methods.

Conclusions:

  • DSCT is a valid and effective method for improving true temporal resolution in coronary CT.
  • DSCT offers significant advantages over EMR in minimizing motion artifacts for clearer cardiac imaging.