Comparison between oscillometric and invasive blood pressure measurements in critically ill premature infants
Sahin Takci1, Sule Yigit, Ayse Korkmaz
1Hacettepe University Ihsan Dogramaci Children's Hospital, Neonatology Unit, Pediatrics, Ankara, Turkey. stakci@gmail.com
Aim:
Blood pressure (BP) measurement techniques in neonates generally involve noninvasive measurements with a cuff (oscillometric) or invasive measurements through an arterial catheter. The aim of this study was to determine the reliability of the noninvasive oscillometric method in critically ill preterm infants when results were compared with the invasive technique.
Method:
Twenty-seven premature infants with a mean birth weight of 1138 ± 552 g were enrolled in the prospective study. Invasive and noninvasive mean arterial pressure (MAP) levels were recorded simultaneously at each measurement in all patients. Low or lower range mean invasive MAP values (MAP ≤30) were evaluated separately as we aimed to assess the value of noninvasive measurements in hypotensive sick premature infants.
Results:
Totally, 431 paired BP measurements were taken during the first week of life. There was no statistically significant difference between invasive and noninvasive readings. However, noninvasive measurements were found significantly higher compared with invasive measurements in the presence of hypotension (p < 0.05).
Conclusion:
This study showed good agreement between oscillometric and invasive readings in critically ill premature infants, and further, comparable mean MAP values were found with the two methods. However, the accuracy of the oscillometric BP measurement technique fails in preterm infants with BP within the lower limits.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer