Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Reliability and the smallest detectable difference of measurements on 3-dimensional cone-beam computed tomography
Janalt Damstra1, Zacharias Fourie, James J R Huddleston Slater
1Department of Orthodontics, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
Introduction:
The aim of this study was to determine the reliability and the measurement error (by means of the smallest detectable error) of 17 commonly used cephalometric measurements made on 3-dimensional (3D) cone-beam computed tomography images.
Methods:
Twenty-five cone-beam computed tomography scans were randomly selected, and 3D images were rendered, segmented, and traced with the SimPlant Ortho Pro software (version 2.1, Materialise Dental, Leuven, Belgium). This was repeated twice by 2 observers during 2 sessions at least 1 week apart. Measurement error was determined by means of the smallest detectable difference. Differences were analyzed with Wilcoxon signed rank tests. Intraobserver and interobserver reliability values were calculated by means of intraclass correlation coefficients (ICC) based on absolute agreement.
Results:
There were great variations of measurement errors between the angular (range, 0.88°-6.29°) and linear (range, 1.33-3.56 mm) variables. The greatest measuring error was associated with the dental measurements U1-FHPL, L1-MdPL. and L1-FHPL (range, 3.80°-6.29°). ANB angle was the only variable with a measuring error of 1° or less for both observers. The intraobserver agreement of all measurements was very good (ICC, 0.86-0.99). Except for SN-FHPL (ICC, 0.76), interobserver agreement was very good (ICC, >0.88).
Conclusions:
The measurement errors of 3D cephalometric measurements (except for the ANB angle) can be considered clinically relevant. This questions the use of linear and angular 3D measurements to detect true treatment effects when a high level of accuracy is required.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...