Related Experiment Video
Updated: May 29, 2026

Leaf Spray Mass Spectrometry: A Rapid Ambient Ionization Technique to Directly Assess Metabolites from Plant Tissues
Published on: June 21, 2018
In situ analysis of agrochemical residues on fruit using ambient ionization on a handheld mass spectrometer
Santosh Soparawalla1, Fatkhulla K Tadjimukhamedov, Joshua S Wiley
1Department of Chemistry, Purdue University, 560 Oval Drive, West Lafayette, IN 47907, USA.
Abstract:
We describe a rapid in situ method for detecting agrochemicals on the surface or in the tissue of fruit using a portable mass spectrometer equipped with an ambient ionization source. Two such ionization methods, low temperature plasma (LTP) and paper spray (PS), were employed in experiments performed at a local grocery store. LTP was used to detect diphenylamine (DPA) directly from the skin of apples in the store and those treated after harvest with DPA were recognized by MS and MS/MS. These data therefore allowed ready distinction between organic and non-organic apples. DPA was also found within the internal tissue of purchased apples and its distribution was mapped using LTP. Similarly, thiabendazole residues were detected on the skin of treated oranges in a grocery store experiment in which paper spray was performed by wiping the orange surface with a moist commercial lens wipe and then applying a high voltage to ionize the chemicals directly from the wipe. The handheld mass spectrometer used in these measurements is capable of performing several stages of tandem mass spectrometry (up to MS(5)); the compounds on the fruit were identified by their MS/MS fragmentation patterns. Protonated DPA (m/z 170) produced a characteristic MS(2) fragment ion at m/z 92, while thiabendazole was identified by MS(3) using precursor to fragment ion transitions m/z 202 →m/z 175 →m/z 131. These particular examples exemplify the power of in situ analysis of complex samples using ambient ionization and handheld mass spectrometers.
More Related Videos
08:10Fizzy Extraction of Volatile Organic Compounds Combined with Atmospheric Pressure Chemical Ionization Quadrupole Mass Spectrometry
Published on: July 14, 2017
05:20Determination of 45 Pesticides in Avocado Varieties by the QuEChERS Method and Gas Chromatography-Tandem Mass Spectrometry
Published on: December 8, 2023
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometers
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Tandem Mass Spectrometry