Endotracheal tube cuff pressure monitoring during neurosurgery - Manual vs. automatic method
Mukul Kumar Jain1, Chander Bushan Tripathi
1Department of Neuroanesthesiology, Institute of Human Behavior and Allied Sciences, Dilshad Garden, Delhi, India.
Background:
Inflation and assessment of the endotracheal tube cuff pressure is often not appreciated as a critical aspect of endotracheal intubation. Appropriate endotracheal tube cuff pressure, endotracheal intubation seals the airway to prevent aspiration and provides for positive-pressure ventilation without air leak.
Materials And Methods:
Correlations between manual methods of assessing the pressure by an experienced anesthesiologists and assessment with maintenance of the pressure within the normal range by the automated pressure controller device were studied in 100 patients divided into two groups. In Group M, endotracheal tube cuff was inflated manually by a trained anesthesiologist and checked for its pressure hourly by cuff pressure monitor till the end of surgery. In Group C, endotracheal tube cuff was inflated by automated cuff pressure controller and pressure was maintained at 25-cm H(2)O throughout the surgeries. Repeated measure ANOVA was applied.
Results:
Repeated measure ANOVA results showed that average of endotracheal tube cuff pressure of 50 patients taken at seven different points is significantly different (F-value: 171.102, P-value: 0.000). Bonferroni correction test shows that average of endotracheal tube cuff pressure in all six groups are significantly different from constant group (P = 0.000). No case of laryngomalacia, tracheomalacia, tracheal stenosis, tracheoesophageal fistula or aspiration pneumonitis was observed.
Conclusions:
Endotracheal tube cuff pressure was significantly high when endotracheal tube cuff was inflated manually. The known complications of high endotracheal tube cuff pressure can be avoided if the cuff pressure controller device is used and manual methods cannot be relied upon for keeping the pressure within the recommended levels.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement