A Novel Application of Musculoskeletal Ultrasound Imaging
Published on: September 17, 2013
Analysis of 2-D motion tracking in ultrasound with dual transducers
Jeffrey M Abeysekera1, Reza Zahiri Azar, Orcun Goksel
1Department of Mechanical Engineering, University of British Columbia, Vancouver, BC, Canada. jeffa@ece.ubc.ca
Abstract:
We study displacement and strain measurement error of dual transducers (two linear arrays, aligned orthogonally and coplanar). Displacements along the beam of each transducer are used to obtain measurements in two-dimensions. Simulations (5MHz) and experiments (10MHz) are compared to measurements with a single linear array, with and without angular compounding. Translation simulations demonstrate factors of 1.07 larger and 8.0 smaller biases in the axial and lateral directions respectively, for dual transducers compared to angular compounding. As the angle between dual transducers decreases from 90° to 40°, for 1% compression simulations, the lateral RMS error ranges from 2.1 to 3.9μm compared to 9μm with angular compounding. Simulation of dual transducer misalignment of 1mm and 2° result in errors of less than 9μm. Experiments demonstrate factors of 3.0 and 5.2 lower biases for dual transducers in the axial and lateral directions respectively compared to angular compounding.