Reduced-force closed trocar entry technique: analysis of trocar insertion force using a mechanical force gauge
James Fanning1, Manish Shah, Bradford Fenton
1Department of Obstetrics and Gynecology, Pennsylvania State University, Milton S. Hershey Medical Center, Hershey, PA 17033, USA. jfanning1@hmc.psu.edu
JSLS : Journal of the Society of Laparoendoscopic Surgeons
|September 10, 2011
Summary
A new reduced-force trocar entry technique significantly lowers insertion force by 50% compared to standard methods. This safe and cost-effective approach may reduce trocar insertion injuries.
Area of Science:
- Minimally Invasive Surgery
- Surgical Techniques
- Patient Safety
Background:
- Trocar insertion is associated with significant patient morbidity, mortality, and healthcare costs.
- Standard trocar entry techniques can exert considerable force, potentially leading to injury.
- Minimizing insertion force is crucial for improving surgical safety and outcomes.
Purpose of the Study:
- To compare the insertion force of a novel reduced-force closed trocar entry technique against the standard closed trocar entry method.
- To quantify the difference in insertion force using a mechanical force gauge.
- To evaluate the safety and feasibility of the reduced-force technique.
Main Methods:
- A prospective trial involving 25 patients was conducted in an operating room setting.
- A mechanical force gauge, integrated into a sterile glove, measured trocar insertion forces.
- Two techniques were compared: standard closed trocar entry and a novel reduced-force closed trocar entry involving pre-dilation with a hemostat.
Main Results:
- The reduced-force trocar entry technique demonstrated a median insertion force of 3.3 lb, significantly lower than the standard technique's 6.5 lb (P=.001).
- The range of forces for the reduced-force technique was 1.6 to 5.4 lb, compared to 2.0 to 14.0 lb for the standard technique.
- No complications were reported with the reduced-force trocar entry technique during the study.
Conclusions:
- The reduced-force closed trocar entry technique effectively decreases trocar insertion force by approximately 50%.
- This technique requires no additional instruments or costs, is rapid, and is safe for patient use.
- Lowering trocar entry force pressure holds potential for reducing the incidence and severity of trocar insertion injuries.
Related Concept Videos
Moment of a Force: Problem Solving
Understanding the scalar formulation of the moment of a force and applying it correctly through problem-solving is crucial in designing and analyzing mechanical systems. Here are the steps for problem-solving with the moment of a force:
Machines: Problem Solving I
A toggle clamp is a mechanical device commonly used for holding and clamping objects in various applications, such as woodworking, metalworking, and assembly operations. Consider a toggle clamp subjected to a force of 200 N at the handle. The vertical clamping force can be calculated, provided the dimensions of the toggle clamp are known.
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...