Related Experiment Video
Updated: May 29, 2026

Testing a Cochlear Implant Electrode Insertion Training System for Optimal Electrode Array Placement in Different Inner Ear Anatomies
Published on: February 6, 2026
Cochlear implant electrode array misplaced in Hyrtl's fissure
Amina Mouzali1, Kheiredidine Ouennoughi, Mohamed-Salah Haraoubia
1Department of Otorhinolaryngology Head & Neck Surgery, Beni Messous Universitary Hospital, Alger 1600, Algeria.
Abstract:
Hyrtl's fissure is a cleft that is present in the developing fetal petrous temporal bone and extends from the area inferior to the round window to the meninges of the posterior fossa. Persistent Hyrtl's fissure, due to incomplete ossification, is considered a rare temporal bone malformation, and is a known cause of perilabyrinthine cerebrospinal fluid fistula. Very few cases are reported as being at risk of complication of cochlear implant surgery. Here we report the case of an 8-year-old boy with misplacement of an electrode array in Hyrtl's fissure. The diagnosis was made postoperatively, since cochlear implant failure was suspected from non-auditory responses. Computed tomography (CT) revealed the extracochlear location of the electrode array. We emphasize the role of presurgical imaging CT and magnetic resonance imaging in detecting temporal bone abnormalities, and we discuss the value of intraoperative auditory nerve response telemetry and postoperative radiological evaluation in diagnosing cochlear implant misplacement.