Related Experiment Video
Updated: May 29, 2026

Preparation of Polyoxometalate-based Photo-responsive Membranes for the Photo-activation of Manganese Oxide Catalysts
Published on: August 7, 2018
New polymeric membrane cadmium(II)-selective electrodes using tripodal amine based ionophores
Utisawadee Khamjumphol1, Sarayut Watchasit, Chomchai Suksai
1Department of Chemistry, Faculty of Science, Khon Kaen University, Khon Kaen, Thailand.
New PVC membrane electrodes show excellent selectivity for cadmium (Cd2+) ions. Ionophore IV, integrated into a plasticized PVC membrane, demonstrated superior performance for cadmium detection in various samples.
Area of Science:
- Electrochemistry
- Materials Science
- Analytical Chemistry
Background:
- Development of selective ion-selective electrodes (ISEs) is crucial for accurate heavy metal detection.
- Cadmium (Cd2+) poses significant environmental and health risks, necessitating sensitive monitoring methods.
Purpose of the Study:
- To design and fabricate novel neutral carrier-based PVC membrane electrodes for selective Cd(2+) detection.
- To investigate the structure-property relationships of synthesized ionophores for optimal electrode performance.
Main Methods:
- Synthesis of ionophores with varying topologies, donor atoms, and lipophilicity, including tripodal amine (TPA) units attached to anthracene and calix[4]arene scaffolds.
- Incorporation of synthesized ionophores into plasticized PVC membranes to create Cd(II) ISEs.
- Optimization of membrane composition (ionic sites, plasticizers) and evaluation of electrode selectivity, response characteristics, and operational stability.
Main Results:
- Ionophore IV, featuring two TPA units on a calix[4]arene core, exhibited the best selectivity for Cd(2+).
- The optimized electrode using ionophore IV demonstrated a Nernstian slope of 29.4±0.6 mV decade(-1), a wide working range (1.6×10(-6)-1.0×10(-2)M), fast response (10s), and stability for over a week.
- The electrodes operated effectively within a pH range of 6.0-9.0 and were successfully applied to determine Cd(2+) in CdS quantum dots solutions and wastewater samples.
Conclusions:
- The developed PVC membrane electrodes based on ionophore IV offer a highly selective and sensitive platform for Cd(2+) determination.
- The study highlights the significant impact of ionophore structure on electrode performance, providing insights for designing advanced chemical sensors.
More Related Videos
08:03Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
Related Concept Videos
Potentiometry: Membrane Electrodes
Ion Exchange
EDTA: Chemistry and Properties
Complexometric Titration: Ligands