Related Experiment Video
Updated: May 29, 2026

Detection of Live Escherichia coli O157:H7 Cells by PMA-qPCR
Published on: February 1, 2014
DuPont qualicon BAX system real-time PCR assay for Escherichia coli O157:H7
Frank Burns1, Lois Fleck, Bridget Andaloro
1DuPont Qualicon, ESL Bldg 400, PO 80400, Rt. 141 and Henry Clay Rd, Wilmington, DE 19880-0400, USA.
Abstract:
Evaluations were conducted to test the performance of the BAX System Real-Time PCR assay, which was certified as Performance Tested Method 031002 for screening E. coli O157:H7 in ground beef, beef trim, spinach, and lettuce. Method comparison studies performed on samples with low-level inoculates showed that the BAX System demonstrates a sensitivity equivalent or superior to the FDA-BAM and the USDA-FSIS culture methods, but with a significantly shorter time to result. Tests to evaluate inclusivity and exclusivity returned no false-negative and no false-positive results on a diverse panel of isolates, and tests for lot-to-lot variability and tablet stability demonstrated consistent performance. Ruggedness studies determined that none of the factors examined affect the performance of the assay. An accelerated shelf life study determined an initial 36 month shelf life for the test kit.