Investigating interferences of a whole-blood point-of-care creatinine analyzer: comparison to plasma enzymatic and
Joely A Straseski1, Martha E Lyon, William Clarke
1University of Utah Health Sciences Center, Department of Pathology, Salt Lake City, USA. joely.a.straseski@aruplab.com
Background:
Although measurement of whole-blood creatinine at the point of care offers rapid assessment of renal function, agreement of point-of-care (POC) results with central laboratory methods continues to be a concern. We assessed the influence of several potential interferents on POC whole-blood creatinine measurements.
Methods:
We compared POC creatinine (Nova StatSensor) measurements with plasma enzymatic (Roche Modular) and isotope dilution mass spectrometry (IDMS) assays in 119 hospital inpatients. We assessed assay interference by hematocrit, pH, pO(2), total and direct bilirubin, creatine, prescribed drugs, diagnosis, red blood cell water fraction, and plasma water fraction.
Results:
CVs for POC creatinine were 1.5- to 6-fold greater than those for plasma methods, in part due to meter-to-meter variation. Regressioncomparison of POC creatinine to IDMS results gave a standard error (S(y|x)) of 0.61 mg/dL (54 μmol/L), whereas regression of plasma enzymatic creatinine to IDMS was S(y|x) 0.16 mg/dL (14 μmol/L). By univariate analysis, bilirubin, creatine, drugs, pO(2), pH,plasma water fraction, and hematocrit were not found to contribute to method differences. However, multivariate analysis revealed that IDMS creatinine, red blood cell and plasma water fractions, and hematocrit explained 91.8% of variance in POC creatinine results.
Conclusions:
These data suggest that whole-blood POC creatinine measurements should be used with caution. Negative interferences observed with these measurements could erroneously suggest adequate renal function near the decision threshold, particularly if estimated glomerular filtration rate is determined. Disparity between whole-blood and plasma matrices partially explains the discordance between whole-blood and plasma creatinine methods.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...