Related Experiment Video
Updated: May 29, 2026

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Exchange of DuraHeart left ventricular assist device via a subcostal approach
Masashi Kawamura1, Taichi Sakaguchi, Shigeru Miyagawa
1Department of Cardiovascular Surgery, Graduate School of Medicine, Osaka University, 2-2E1 Yamadaoka, Suita, Osaka 565-0871, Japan.
Abstract:
We report a successful case of a DuraHeart left ventricular assist device (LVAD) exchange via a subcostal approach. A 35-year-old woman was implanted with a DuraHeart LVAD due to dilated cardiomyopathy. Approximately 8 months after the implantation, the magnetic levitation system failed. The DuraHeart LVAD was exchanged emergently. The pump was freely dissected via a subcostal approach, avoiding redo sternotomy. De-airing of the new pump and the left ventricle was carefully performed. When the systemic flow was transferred from the cardiopulmonary bypass to the DuraHeart LVAD, an adequate flow was not initally obtained. Positional correction of the inflow conduit was needed to obtain full systemic flow. The postoperative course was uneventful. She was successfully discharged and is waiting at home for a heart donation.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization