Indirect computed tomography venography with a low-tube-voltage technique: reduction in the radiation and contrast
Seitaro Oda1, Daisuke Utsunomiya, Kazuo Awai
1Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan. seisei0430@nifty.com
Objective:
The objective of the study was to investigate the effects of low-tube-voltage computed tomography (CT) venography on qualitative and quantitative image parameters and the radiation dose.
Methods:
Eighty-eight studies on 84 patients underwent pelvic and lower-extremity CT venography under protocol A (standard 120 kV with 150 mL of contrast material, n = 44) or protocol B (80 kV with 100 mL of contrast material, n = 44) on a 64-detector CT scanner. We compared the dose length product in the 2 protocols. Two blinded observers measured CT attenuation in the veins, the image noise, contrast-to-noise ratio, and figure of merit.
Results:
The mean dose length product was significantly lower under protocol B than A (603.2 [SD, 67.2] vs 1131.7 [SD, 67.0] mGy × cm) (P < 0.01). Mean CT attenuation of the veins was significantly greater with protocol B (125.3 [SD, 16.2] vs 106.1 [SD, 16.0] Hounsfield units) (P < 0.01), and the mean image noise was also significantly higher under protocol B (6.6 [SD, 0.8] vs 4.9 [SD, 0.7] Hounsfield units) (P < 0.01). There was no difference in contrast-to-noise ratio (P = 0.46). Figure of merit was significantly higher under protocol B (P < 0.01).
Conclusions:
Computed tomography venography with a low-tube-voltage technique allows reducing the radiation dose and the amount of contrast material without image quality degradation.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging