Related Experiment Video
Updated: May 29, 2026

05:31
Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
The optional VenaTech(™) Convertible (™) vena cava filter: experimental study in sheep.
Alain F Le Blanche1, Jean-Baptiste Ricco, Michel Bonneau
1University of Picardie-Jules Verne, Amiens, France. le_blanche.alain@chu-amiens.fr
Cardiovascular and Interventional Radiology
|September 20, 2011
Summary
This study shows the VenaTech Convertible inferior vena cava (IVC) filter is safe and effective for retrieval. The convertible IVC filter demonstrated successful implantation and conversion, with no migration or tilting observed.
Area of Science:
- Vascular Surgery
- Interventional Radiology
- Biomedical Engineering
Background:
- Retrieval of optional inferior vena cava (IVC) filters can be challenging due to complications like tilting, migration, fracture, or embedding.
- A novel convertible IVC filter designed for easier retrieval by removing the filter head was developed.
Purpose of the Study:
- To evaluate the safety and feasibility of a new convertible IVC filter.
- To assess the performance of the VenaTech Convertible IVC filter in terms of implantation, conversion, and retrieval.
Main Methods:
- Forty-nine sheep underwent IVC catheterization and insertion of VenaTech Convertible IVC filters.
- Filters were either permanently placed or designated for conversion immediately or after 1, 3, or 6 months.
- Filter performance, including migration, tilting, and IVC wall incorporation, was assessed via cavography and gross anatomy.
Main Results:
- Successful delivery, deployment, and immediate conversion were achieved in all cases.
- Delayed conversion was successful in all but one animal, with 12 requiring balloon-catheter assistance.
- No thrombosis, migration, or significant tilting was observed in any of the 49 sheep. Filter components incorporated into the IVC wall within 4 weeks, with no intimal growth on the removed head.
Conclusions:
- The VenaTech Convertible IVC filter demonstrated successful implantation in all subjects.
- Filter conversion via head removal was feasible at various time points, indicating potential for easier retrieval.
- The study supports the safety and efficacy of this convertible IVC filter design.