Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Three-dimensional CT vascular mapping of a juxtauterine pelvic schwannoma suggesting retroperitoneal origin: a case report.

Case reports in women's health·2026
Same author

Investigation of a storage method for peritoneal and pleural effusion cytology specimens for long-term preservation of antigenicity in immunocytochemical staining.

Acta cytologica·2026
Same author

Fatal Intra-abdominal Hemorrhage in Eosinophilic Granulomatosis with Polyangiitis: An Autopsy Case Report.

Internal medicine (Tokyo, Japan)·2026
Same author

Pitfalls in the differential diagnosis of diarrhea after kidney transplantation: challenges in identifying Yersinia enterocolitica infection.

CEN case reports·2026
Same author

Carcinosarcoma in Ovarian Mature Teratoma: A Rare Autopsy Case.

Cureus·2025
Same author

Hematometra within a cesarean scar defect in a perimenopausal woman: A case report.

The journal of obstetrics and gynaecology research·2025

Related Experiment Video

Updated: May 29, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

[Optimization of exposure conditions for amorphous selenium direct conversion DR-based mammography system].

Tokiko Endo1, Misaki Shiraiwa, Mikinao Oiwa

  • 1Department of Advanced Diagnosis, Clinical Research Center, National Hospital Organization Nagoya Medical Center.

Nihon Hoshasen Gijutsu Gakkai Zasshi
|September 23, 2011
PubMed
Summary

Optimizing digital radiography (DR) mammography exposure conditions enhances microcalcification and mass detection. Different settings are best for phantom analysis versus actual breast tissue imaging, improving diagnostic accuracy.

More Related Videos

Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
08:46

Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages

Published on: April 13, 2016

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

Published on: September 14, 2018

Related Experiment Videos

Last Updated: May 29, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages
08:46

Using Synchrotron Radiation Microtomography to Investigate Multi-scale Three-dimensional Microelectronic Packages

Published on: April 13, 2016

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

Published on: September 14, 2018

Area of Science:

  • Medical Imaging
  • Radiologic Technology
  • Diagnostic Radiology

Background:

  • Digital radiography (DR) mammography offers improved visualization of breast lesions.
  • Optimizing exposure parameters is crucial for enhancing image quality and diagnostic accuracy in mammography.

Purpose of the Study:

  • To determine optimal exposure conditions (target/filter, tube voltage) for DR mammography.
  • To compare phantom-based physical evaluations with visual assessments of breast specimens for optimizing image quality.
  • To improve the detectability of microcalcifications and masses in mammographic images.

Main Methods:

  • Evaluated contrast-to-noise ratios (CNRs) in polymethyl methacrylate (PMMA) phantoms of varying thicknesses under consistent average glandular doses (AGDs).
  • Compared physical measurements with visual evaluations of 50 breast specimens irradiated under diverse exposure conditions.
  • Controlled image histograms for consistency during visual assessment.

Main Results:

  • Tungsten/rhodium targets/filters yielded the highest CNRs in phantoms ≥20 mm thick.
  • Molybdenum/molybdenum was optimal for visualizing microcalcifications and masses in specimens ≤35 mm thick.
  • Molybdenum/rhodium and tungsten/rhodium provided superior image quality for specimens >35 mm and >55 mm, respectively, at consistent AGDs.

Conclusions:

  • Optimal exposure conditions for maximizing phantom CNR differ from those yielding the best diagnostic image quality for breast specimens.
  • Visual evaluation of breast specimens is essential for optimizing DR mammography exposure settings for accurate and efficient clinical diagnosis.
  • The findings enable more precise and effective mammographic examinations.