Related Experiment Video
Updated: May 29, 2026

Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
The effect of CT dose on glenohumeral joint congruency measurements using 3D reconstructed patient-specific bone
Emily A Lalone1, Anne-Marie V Fox, Angela E Kedgley
1Biomedical Engineering, The University of Western Ontario, London, Canada.
Abstract:
The study of joint congruency at the glenohumeral joint of the shoulder using computed tomography (CT) and three-dimensional (3D) reconstructions of joint surfaces is an area of significant clinical interest. However, ionizing radiation delivered to patients during CT examinations is much higher than other types of radiological imaging. The shoulder represents a significant challenge for this modality as it is adjacent to the thyroid gland and breast tissue. The objective of this study was to determine the optimal CT scanning techniques that would minimize radiation dose while accurately quantifying joint congruency of the shoulder. The results suggest that only one-tenth of the standard applied total current (mA) and a pitch ratio of 1.375:1 was necessary to produce joint congruency values consistent with that of the higher dose scans. Using the CT scanning techniques examined in this study, the effective dose applied to the shoulder to quantify joint congruency was reduced by 88.9% compared to standard clinical CT imaging techniques.

