Comparison of invasive and non-invasive measurements of haemodynamic parameters in patients with advanced heart

Mateusz Sokolski1, Agnieszka Rydlewska, Bartosz Krakowiak

  • 1Department of Heart Diseases, Wroclaw Medical University, Wroclaw, Poland. matsok@gmail.com

Insights

A non-invasive method for measuring hemodynamic parameters like stroke volume, cardiac output, and systemic vascular resistance shows adequate concordance with the invasive Swan-Ganz catheter in advanced heart failure patients.

Area of Science:

  • Cardiology
  • Medical Devices
  • Hemodynamics

Background:

  • Hemodynamic monitoring is crucial for advanced heart failure management.
  • Invasive methods like Swan-Ganz catheters have limitations due to their invasiveness.
  • Non-invasive alternatives are sought to improve patient care and monitoring accessibility.

Purpose of the Study:

  • To assess the agreement between invasive (Swan-Ganz catheter) and non-invasive hemodynamic measurements.
  • To evaluate the clinical applicability of a novel non-invasive device for hemodynamic monitoring in advanced heart failure.
  • To determine the concordance of stroke volume, cardiac output, and systemic vascular resistance measurements using both methods.

Main Methods:

  • Simultaneous measurement of resting hemodynamic parameters in 25 advanced heart failure patients.
  • Comparison between invasive Swan-Ganz catheterization and a non-invasive finger arterial pressure waveform analysis device.
  • Assessment of concordance using the Bland-Altman method and calculation of variability coefficients for stroke volume, cardiac output, and systemic vascular resistance.

Main Results:

  • Variability coefficients were 13% for stroke volume, 18% for systemic vascular resistance, and 11% for cardiac output.
  • Adequate concordance was observed between invasive and non-invasive measurements across various patient subgroups.
  • The non-invasive method demonstrated consistent results regardless of decompensation status or atrial fibrillation.

Conclusions:

  • Non-invasive hemodynamic monitoring using a pressure pulse contour model shows sufficient agreement with invasive methods.
  • This non-invasive approach holds promise for routine clinical application in advanced heart failure patients.
  • The findings support the potential of non-invasive techniques to enhance hemodynamic monitoring in this population.