A randomized trial of the Vein Viewer versus standard technique for placement of peripherally inserted central
K Phipps1, A Modic, M A O'Riordan
1Division of Neonatology, Department of Pediatrics, Rainbow Babies and Children's Hospital, University Hospitals of Cleveland Case Medical Center, Cleveland, OH, USA. karla.phipps@UHhospitals.org
Objective:
Peripherally inserted central catheters are important but can be difficult to place in neonates. Therefore, we compared a near-infrared device, the Vein Viewer, to determine if its use would increase successful line placement, with standard techniques.
Study Design:
Randomized controlled trial in preterm and term neonates in a level 3 Neonatal Intensive Care Unit.
Result:
In all, 115 subjects were enrolled with 59 randomized to the Vein Viewer group and 56 to the control group. Overall, use of the Vein Viewer showed a trend to more successful placement 86 versus 75%; unadjusted odds ratio 2.33 (0.90, 6.04; P=0.08). Infants randomized to the Vein Viewer were more mature (30 ± 2 weeks gestational age (GA) versus 28 ± 2 weeks GA; P=0.08). After adjusting for GA, use of the Vein Viewer was significantly more likely to lead to successful line placement (adjusted odds ratio 3.05 (1.10, 1.82)).
Conclusion:
The Vein Viewer improved successful placement with the most benefit seen in infants of greater GA.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias