The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Comparison of the effects of two humidifier systems on endotracheal tube resistance
Indalecio Morán1, Belén Cabello, Elizabeth Manero
1Servei Medicina Intensiva, Hospital de la Santa Creu i Sant Pau, Universitat Autònoma de Barcelona, Sant Quintí, 89, 08041, Barcelona, Spain.
Purpose:
To compare the effects of two humidifier systems on endotracheal tube (ETT) resistance during mechanical ventilation, either an active heated humidifier (HH) or a passive heat and moisture exchanger (HME) was selected using current clinical recommendations.
Methods:
This was a prospective clinical cohort study performed in an intensive care unit. Gas conditioning was performed using the HH in 22 patients and the HME in another 22. Patients were matched for endotracheal tube diameter, days of mechanical ventilation, simplified acute physiology score II (SAPS II), and fluid balance.
Results:
Used-ETT resistance was measured immediately after extubation. Unused-ETT resistance was calculated with an identical, clean ETT. No differences were found between the HH and HME groups in ETT diameter (7.9 ± 0.4 vs. 7.9 ± 0.3 mm; p = 0.98), days of mechanical ventilation (11.3 ± 7.7 vs. 9.5 ± 4.5; p = 0.34), SAPS II (41.0 ± 13.6 vs. 42.0 ± 11.7; p = 0.79), or fluid balance (-2,552 ± 6,268 vs. -2,579 ± 5,422 mL; p = 0.98). ETT resistance increased from intubation to extubation: from 6.8 ± 1.1 to 10.6 ± 4.3 cmH(2)O L(-1) s(-1) in the HH group, (p < 0.001) and from 6.8 ± 1.1 to 10.2 ± 3.8 cmH(2)O L(-1) s(-1) in the HME group (p < 0.001), which is a 53% average increase in resistive load.
Conclusions:
We did not find differences between the two types of humidifiers in terms of airflow resistance during prolonged mechanical ventilation when the devices were selected on the basis of individual clinical needs. The increase in resistive load is physiologically relevant.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...