Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
What does built-in software of home ventilators tell us? An observational study of 150 patients on home ventilation
Patrick Pasquina1, Dan Adler, Pamela Farr
1Division of Pulmonary Diseases, Geneva University Hospital, Geneva, Switzerland.
Background:
Recent home ventilators are equipped with built-in software which provides data such as compliance, estimations of leaks, tidal volume, minute ventilation, respiratory rate, apnea and apnea-hypopnea indexes, and percentage of inspirations triggered by the patient (or ventilator). However, for many of these variables, there is neither consensus nor documentation as to what is to be expected in a population of stable patients under noninvasive ventilation (NIV).
Objectives:
To document the values and distribution of specific items downloaded from ventilator monitoring software, by diagnostic category.
Methods:
Analysis of data downloaded from home ventilators in clinically stable patients under long-term NIV, during elective home visits by specialized nurses.
Results:
Data were collected from home ventilators of 150 patients with chronic obstructive pulmonary disease (n = 32), overlap syndrome (n = 29), obesity-hypoventilation (n = 38), neuromuscular disorders (n = 19), restrictive disorders (n = 21), and central sleep apnea syndrome (n = 11). On average, leaks were low, being lowest in patients with facial masks (vs. nasal masks), and increased with older age. Compliance was excellent in all groups. Patients with neuromuscular diseases triggered their ventilators less and tended to be 'captured', while other groups triggered at least half of inspiratory cycles. Most patients had a respiratory rate just slightly above the back-up rate. Residual apneas and hypopneas were highest in patients with central apneas.
Conclusions:
Built-in software of home ventilators provides the clinician with new parameters, some of which are a useful adjunct to recommended tools for monitoring NIV and may contribute to a better understanding of residual hypoventilation and/or desaturations. However, an independent validation of the accuracy of this information is mandatory.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...