Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study
1Centre Hospitalier et Universitaire de Poitiers, Université de Poitiers et Inserm ERI23, France. o.mimoz@chu-poitiers.fr
Background:
The aim of the study was to assess the reliability of hemoglobin measurements made with HemoCue®, compared with those made with the reference method in critically ill patients.
Methods:
A prospective observational study was conducted in three adult surgical intensive care units of a university hospital. One hundred and ninety-eight consecutive patients were included, and a total of 1166 hemoglobin concentrations were measured using arterial blood samples in the laboratory (HbLAB) and at bedside (HbHC.art) with a portable hemoglobinometer (HemoCue® Hb201+). Simultaneously, a capillary measurement (HbHC.cap) was performed at bedside using the same device.
Results:
The mean difference (bias) between HbHC.cap and HbLAB was 0.2 g/dL (95%CI, 0.1;0.3), and limits of agreement were -1.3 g/dL (95%CI, -1.4;-1.2) to 1.7 g/dL (95%CI, 1.6;1.9). The discrepancies between HbHC.cap and HbLAB were greater than 1 g/dL in 30.8% of cases. The bias between HbHC.art and HbLAB was -0.1 g/dL (95% CI, -0.2;0.2), and limits of agreement were slightly better at -1.1 g/dL (95% CI, -1.2;-1.0) and 1.0 g/dL (95% CI, 0.9;1.1). The HemoCue®'s accuracy was not affected by the hospital unit, the puncture site (finger or ear), norepinephrine administration or by hemoglobin levels below 10 g/dL or 8 g/dL.
Conclusion:
Capillary HemoCue® is not sufficiently accurate to make a therapeutic decision such as whether a blood transfusion should be performed. The method's performance was moderately improved by the use of arterial blood.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...