Dermatophyte identification using matrix-assisted laser desorption ionization-time of flight mass spectrometry
Elitza S Theel1, Leslie Hall, Jayawant Mandrekar
1Division of Clinical Microbiology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Hilton 868, 200 First Street SW, Rochester, MN 55905, USA.
Abstract:
The performance of the Bruker Biotyper matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometer (MS) for the identification of dermatophytes from clinical cultures was compared to that of dermatophyte identification using 28S rRNA gene sequencing. The MALDI Biotyper library (MBL; version 3.0) was used alone and in combination with a supplemented library containing an additional 20 dermatophyte spectra (S-MBL). Acquired spectra were interpreted using both the manufacturer-recommended scores (genus, ≥1.7; species, ≥2.0) and adjusted cutoff values established by this study (genus, ≥1.5; species, ≥1.7); identifications required a minimum 10% difference in scores between the top two different organisms to be considered correct. One hundred well-characterized, archived dermatophyte isolates and 71 fresh dermatophyte cultures were evaluated using both libraries and both sets of cutoff criteria. Collectively, the S-MBL significantly outperformed the MBL at both the genus (93% versus 37.4%; P < 0,0001) and species (59.6% versus 20.5%; P < 0.0001) levels when using the adjusted score criteria. Importantly, application of the lowered cutoff values significantly improved genus (P = 0.005)- and species (P < 0.0001)-level identification for the S-MBL, without leading to an increase in misidentifications. MALDI-TOF MS is a cost-effective and rapid alternative to traditional or molecular methods for dermatophyte identification, provided that the reference library is supplemented to sufficiently encompass clinically relevant, intraspecies strain diversity.
More Related Videos
07:49Matrix-assisted Laser Desorption/Ionization Time of Flight (MALDI-TOF) Mass Spectrometric Analysis of Intact Proteins Larger than 100 kDa
Published on: September 9, 2013
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight (MALDI-TOF) Mass Spectrometry
Published on: June 10, 2018
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
MALDI-TOF Mass Spectrometry
Rapid Identification of Pathogens
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrum: Interpretation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...