Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
Published on: November 7, 2014
Concurrent validity of the Polar s3 Stride Sensor for measuring walking stride velocity
Nicole Grigg1, James Smeathers, Scott Wearing
1Institute of Health and Biomedical Innovation, Queensland University of Technology, Kelvin Grove, QLD, Australia. nicole.grigg@qut.edu.au
Abstract:
With this research, we sought to establish the accuracy of stride velocity data collected by the s3 Stride Sensor Participants walked along a GAITRite mat at self-selected slow, preferred, and fast velocities, with two s3 Stride Sensors attached to their right foot. The start position was systematically varied such that the GAITRite system would record the second through sixth strides at each walking velocity. Both slow and preferred walking velocities were underestimated by 14% relative to the GAITRite (p < .05), while independent of walking velocity, Strides 2 and 3 were underestimated by 26% and 9% (p < .05), respectively. Researchers should use caution when interpreting data collected at slow and preferred walking velocities and during the first three strides.