Related Experiment Video
Updated: May 28, 2026

Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Wheelchair-mounted accelerometers for measurement of physical activity
Shawn C Gendle1, Mark Richardson, James Leeper
1Department of Kinesiology, The University of Alabama, Tuscaloosa, AL 35401, USA. sgendle@cba.ua.edu
Purpose:
To evaluate the validity of a wheelchair frame-mounted accelerometer for the assessment of physical activity of wheelchair users.
Method:
Twelve collegiate wheelchair basketball players participated in this study. The study was conducted in a modern indoor gymnasium at a university in the USA. A randomized, crossover experimental design was used to investigate accelerometer output, participant heart rate, and distance travelled. Participants performed two trials of wheeling at a combination of two different effort levels (light and moderate: Prescribed using perceived exertion) and two different modes (continuous and stop-go).
Results:
Accelerometer vector magnitude activity counts (VM), heart rate (HR), and distance travelled were significantly different between light and moderate effort (p < 0.01). The continuous and stop-go wheeling modes were not significantly different (p = 0.37) for VM, but were significantly different (p < 0.01) for both HR and distance. Between-trial reliability of all data collapsed for the effort and mode combinations were: VM: r = 0.85; HR: r = 0.86; and distance: r = 0.96.
Conclusion:
A wheelchair frame-mounted accelerometer differentiated between perceptually-prescribed low and moderate effort levels and may prove to be a valid instrument in the detection of a wheelchair users' physical activity. [Box: see text].