Related Experiment Video
Updated: Jan 12, 2026

Non-restraining EEG Radiotelemetry: Epidural and Deep Intracerebral Stereotaxic EEG Electrode Placement
Published on: June 25, 2016
Problem of signal contamination in interhemispheric dual-sided subdural electrodes
George Nune1, Jonathan Winawer, Andreas M Rauschecker
1Laboratory of Behavioral & Cognitive Neurology, Department of Neurology & Neurological Sciences, Stanford University, Stanford, California 94305, USA.
Abstract:
Dual-sided subdural electrodes are used in the localization and lateralization of seizure-onset zones when the area of interest is within the interhemispheric fissure. We designed the current study to test the validity of the assumption that each side of the dual-sided electrodes records from the hemisphere it faces. We recorded with dual-sided strip and grid electrodes implanted in the occipital interhemispheric space in two patients with nonoccipital epilepsy during two visual stimulation tasks in which subjects were presented with visual stimuli in the ipsilateral or contralateral visual hemifields. Our findings show substantial contamination of recordings from the opposite hemisphere. Although, as expected, electrodes recording through the falx record faintly from the contralateral cortical surface, they unexpectedly pick up strong signals from the cortex behind them. Therefore, we conclude that these electrodes should not be used for lateralization of the origin of epileptic activity or evoked responses.