High energy resolution bandpass photon detector for inverse photoemission spectroscopy
M Maniraj1, S W D'Souza, J Nayak
1UGC-DAE Consortium for Scientific Research, Khandwa Road, Indore, 452001, Madhya Pradesh, India.
Abstract:
We report a bandpass ultraviolet photon detector for inverse photoemission spectroscopy with energy resolution of 82 ± 2 meV. The detector (Sr(0.7)Ca(0.3)F(2)/acetone) consists of Sr(0.7)Ca(0.3)F(2) entrance window with energy transmission cutoff of 9.85 eV and acetone as detection gas with 9.7 eV photoionization threshold. The response function of the detector, measured using synchrotron radiation, has a nearly Gaussian shape. The n = 1 image potential state of Cu(100) and the Fermi edge of silver have been measured to demonstrate the improvement in resolution compared to the CaF(2)/acetone detector. To show the advantage of improved resolution of the Sr(0.7)Ca(0.3)F(2)/acetone detector, the metal to semiconductor transition in Sn has been studied. The pseudogap in the semiconducting phase of Sn could be identified, which is not possible with the CaF(2)/acetone detector because of its worse resolution.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Instrumentation
Gas Chromatography: Types of Detectors-II
IR Spectrometers
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...