Related Experiment Video
Updated: May 28, 2026

Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Bilateral patellar component shear failure of highly cross-linked polyethylene components: report of a case and
Bernard N Stulberg1, Timothy M Wright, Alex P Stoller
1Orthopaedic and Rheumatologic Institute, Cleveland Clinic, Cleveland, Ohio 44113, USA.
Abstract:
A case of bilateral patellar component failure due to fatigue fracture of the all-polyethylene fixation pegs in a highly cross-linked ultra-high-molecular-weight polyethylene design is presented. To recreate this failure mode, a novel test method was developed to investigate the effects of peg orientation and cement technique on patella fatigue strength under cyclic compression and shear loading. Patellar peg orientation had a minor effect on shear strength, whereas lack of cement in the backside patellar groove had a substantial effect. The shear fatigue strength exceeded in vivo force estimates when the patellar groove was fully cemented. The test results and retrieval analysis suggest that high activity level and inadequate cement fixation of the patellar component may contribute to all-polyethylene patellar component peg fractures.