Micro-CT for anatomic referencing in PET and SPECT: radiation dose, biologic damage, and image quality
Veerle Kersemans1, James Thompson, Bart Cornelissen
1Department of Oncology, CRUK/MRC Gray Institute for Radiation Oncology and Biology, University of Oxford, Oxford, UK. Veerle.Kersemans@oncology.ox.ac.uk
Unlabelled:
CT is widely used for anatomic referencing of PET and SPECT images of small animals but requires sufficiently high radiation doses capable of causing significant DNA damage. Therefore, we described the relationship between radiation dose, biologic damage, and image quality to determine whether CT can be used without significantly compromising radiotherapy and tumor development studies.
Methods:
The CT dose index generated by the nanoSPECT/CT system was compared with measurements using EBT2 gafchromic film. The effects of micro-CT were evaluated in 2 mouse strains that differ in sensitivity to radiation. γH2AX foci analysis to determine leukocyte, liver, and jejunum DNA damage and hematoxylin and eosin staining to investigate macroscopic jejunum damage were performed. Signal-to-noise ratio, contrast-to-noise ratio, and scanner linearity were determined to assess image quality.
Results:
For the standard settings, that is, as set by the manufacturers, EBT2 gafchromic film dosimetry showed that the nanoSPECT/CT system underestimated the absorbed dose. Moreover, significant doses were obtained, resulting in a significant increase in γH2AX formation in leukocytes, liver, and jejunum 40 min after CT, using preset parameters when compared with nonimaged controls. The jejenum response was more pronounced for the more radiosensitive strain. In contrast to leukocytes, the liver and jejunum still showed evidence of DNA damage 3 d after CT. Contrast-to-noise ratio, signal-to-noise ratio, and scanner linearity were sufficient to allow for anatomic referencing for both imaging protocols tested.
Conclusion:
Anatomic reference images can be produced with no observable DNA damage or compromising image quality using low radiographic voltage, flux, and duration.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET