The effect of sterilisation on the light intensity of the reusable Trachlight™
A D Milne1, A L Cave2, P A Brousseau3
1Assistant Professor, Staff Anaesthesiologist, Department of Anaesthesia and School of Biomedical Engineering, Dalhousie University, Halifax, Canada.
Abstract:
We examined the light intensity delivered by the Trachlight™ reusable lightwand and the effect of repeated sterilisation on this light intensity. Using a light meter and test chamber, the light intensity of 12 new Trachlights and 31 Trachlights available in operating theatres was measured. In addition, five new devices underwent 30 repeated sterilisations with light intensity measurement after each cleaning. There was no difference between the light intensity of new Trachlights (median (IQR [range]) 2282 (2202-2780 [1970-4400])) lux and the devices currently in service (2295 (2018-2796 [1649-3280]) lux (p = 0.27)). The median (range) number of uses for devices currently in service was only 3 (0-9). There was, however, significant light intensity loss after 10, 20 and 30 cleanings (p = 0.001). After 10 cleanings, the mean light intensity was 75% of initial output, and after 30 cycles, light intensity was 59% of initial output.
More Related Videos
08:56Investigation of a Blue Light LED Device to Suppress Wound Pathogens Using a Collagen-Based Synthetic Skin Model
Published on: February 24, 2026
08:16Method for the Assessment of Effects of a Range of Wavelengths and Intensities of Red/near-infrared Light Therapy on Oxidative Stress In Vitro
Published on: March 21, 2015
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: