X-ray attenuation coefficient of mixtures: inputs for dual-energy CT
Rezvan Ravanfar Haghighi1, S Chatterjee, Akondi Vyas
1All India Institute of Medical Sciences, Ansari Nagar, New Delhi, India. chat@iiap.res.in
Purpose:
The attenuation coefficient, μ(E) of substances, at any energy (E) of the x-ray photon, is known to depend on the electron density (ρ(e)) and the effective atomic number (Z(eff)) of the material. While the dependence on ρ(e) is known to be linear, that of Z(eff) is found to follow a power law (Z(eff))(x) which makes it very sensitive to the index "x". Several different values, lying between 3 and 4 have been suggested for the exponent x, in the existing literature. The purpose of the present investigations is to ascertain empirically the value that should be assigned to x.
Methods:
This is done by measuring the HU value of different mixtures, having different values for ρ(e) and Z(eff) (calculated from their known chemical compositions) and thus determining the dependence of their attenuation coefficients (μ) on the above two quantities.
Results:
The experimental results show the dependence of μ on Z(eff) to be of the power law type, [ρ(e)(Z(eff))(x)/E(y)], where y = 3.0669 but no single value for the index x, can be assigned to fit the observed data. It is seen in different mixtures that the value of x predominantly decreases as Z(eff) increases from 7.5 to 12.
Conclusions:
This result points out that very large errors can occur in calculating Z(eff) from the values of μ if a fixed value for x is used. The importance of this result to dual energy computed tomography is pointed out and it is concluded that the proper values for x are required to be incorporated in the inversion algorithms, for the different regimes of Z(eff).
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...