A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
First prospective, multi-centre clinical experience with a novel left ventricular quadripolar lead
Johannes Sperzel1, Wilfried Dänschel, Klaus-Jürgen Gutleben
1Cardiology Department, Kerckhoff-Klinik GmbH, Benekestr 2-8, 61231, Bad Nauheim, Germany. j.sperzel@kerckhoff-klinik.de
A new quadripolar electrode lead for cardiac resynchronization therapy (CRT) offers more programming options to manage high pacing thresholds and phrenic nerve stimulation (PNS), potentially avoiding lead repositioning.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Cardiac resynchronization therapy (CRT) can be complicated by elevated pacing thresholds and phrenic nerve stimulation (PNS).
- These complications often necessitate coronary sinus lead repositioning.
- A novel quadripolar electrode lead and CRT-defibrillator (CRT-D) system aims to mitigate these issues through electrical repositioning.
Purpose of the Study:
- To evaluate the performance of a novel quadripolar electrode lead and CRT-D device.
- To assess the potential of electrical repositioning to avoid lead revision procedures.
- To determine the efficacy in managing pacing thresholds and PNS.
Main Methods:
- Seventy-five patients indicated for CRT were enrolled and implanted with a quadripolar electrode lead and CRT-D device.
- Electrical data and presence of PNS were documented for each left ventricular (LV) configuration.
- Measurements were taken pre-hospital discharge and at 1-month follow-up.
Main Results:
- 71 patients were successfully implanted with the quadripolar lead.
- Electrical measurements remained stable throughout the follow-up period.
- 97% of patients had programmable configurations with < 2.5 V threshold and no PNS, compared to 86% with conventional bipolar configurations.
- Physicians utilized programming options to manage threshold changes and PNS effectively.
- 11% of patients experienced PNS on all conventional bipolar vectors.
Conclusions:
- The quadripolar electrode LV lead offers enhanced programming flexibility for CRT management.
- New electrical vector measurements are comparable to conventional configurations.
- This technology provides a valuable alternative for addressing common CRT complications like PNS.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Catheterization I: Pre-Procedure Overview